Agilent InfinityLab LC Series RID User Manual 108
8 Test Functions
Refractive Index Calibration
Refractive Index Calibration
Refractive Index Calibration
The refractive index calibration is based on a sucrose calibration solution, which
has a known refractive index compared to LC grade water. After both the sample
and reference cells have been purged with LC grade water the sucrose solution is
introduced into the flow cell and then the built-in refractive index calibration
functionality is used.
Filling the sample cell with the sucrose calibration solution will give a theoretical
detector response of 512,000 nRIU +/- 5,000 nRIU. The calibration algorithm will
allow the actual detector response, if different, to be changed to the theoretical
value.
NOTE
Refractive index calibration is only required after exchange of the optical unit or
the main (RIM) - board.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals