Agilent InfinityLab LC Series RID User Manual 97
7Error Information
Refractive Index Detector Specific Error Messages
Purge Valve Fuse Blown
Error ID: 2686
The electronic fuse on the purge valve has been activated.
Recycle Valve Fuse Blown
Error ID: 2687
The electronic fuse on the recycle valve has been activated.
Probable cause Suggested actions
1
Short in purge valve circuit.
Power cycle the module.
2
Defective purge valve.
Please contact your Agilent service representa-
tive.
3
Defective main board.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Short in recycle valve circuit.
Power cycle the module.
2
Defective recycle valve.
Please contact your Agilent service representa-
tive.
3
Defective main board.
Please contact your Agilent service representa-
tive.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals