Agilent InfinityLab LC Series RID User Manual 191
13 LAN Configuration
Manual Configuration
Manual Configuration
Manual configuration only alters the set of parameters stored in the non-volatile
memory of the module. It never affects the currently active parameters.
Therefore, manual configuration can be done at any time. A power cycle is
mandatory to make the stored parameters become the active parameters, given
that the initialization mode selection switches are allowing it.
Figure 38 Manual Configuration (Principle)
TELNET
Session
Control
Module
Stored
Parameter
Non-Volatile
RAM