Agilent InfinityLab LC Series RID User Manual 115
8 Test Functions
Optical Balance
The Optical Balance Procedure
1
Run the RID Tools Screen with the Agilent Lab Advisor (for further information
see Online-Help of user interface). (If an Agilent pump is part of the system,
the pump section is active.)
2
Purging the sample and reference cells.
a
Switch the purge valve to the
ON
position.
b
Purge the sample and reference cells for around 10 min with the solvents
to be used.
c
Switch the purge valve to the
OFF
position
3
Select
Optical Balance
.
When
When light falling on light receiving diodes is out of balance.
Tools required Description
Flat head screwdriver
NOTE
This procedure should only be performed to correct a permanent misalignment
of the light beam that cannot be eliminated by flushing the sample and the
reference cell with the same solvent and by equilibrating the system.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals