Agilent InfinityLab LC Series RID User Manual 99
7Error Information
Refractive Index Detector Specific Error Messages
Lamp Voltage too Low
Error ID: 2693
Lamp Voltage too High
Error ID: 2690
Lamp Current too High
Error ID: 2691
Probable cause Suggested actions
1
Defective main board.
Please contact your Agilent service representa-
tive.
2
Defective lamp or optics.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Contaminated flow cell.
Flush the flow cell.
2
Defective main board.
Please contact your Agilent service representa-
tive.
3
Defective lamp or optics.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Defective main board.
Please contact your Agilent service representa-
tive.
2
Defective lamp or optics.
Please contact your Agilent service representa-
tive.