7Error Information
Refractive Index Detector Specific Error Messages
Lamp Voltage too Low
Error ID: 2693
Lamp Voltage too High
Error ID: 2690
Lamp Current too High
Error ID: 2691
Probable cause Suggested actions
1
Defective main board.
Please contact your Agilent service representa-
tive.
2
Defective lamp or optics.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Contaminated flow cell.
Flush the flow cell.
2
Defective main board.
Please contact your Agilent service representa-
tive.
3
Defective lamp or optics.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Defective main board.
Please contact your Agilent service representa-
tive.
2
Defective lamp or optics.
Please contact your Agilent service representa-
tive.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals