Agilent InfinityLab LC Series RID User Manual 119
8 Test Functions
ASTM Drift and Noise Test
ASTM Drift and Noise Test
Setting the Test Conditions
1
Turn
ON
the detector.
You are now ready to change the settings of your detector.
2
Connect the restriction capillary directly between the column compartment
heat exchanger outlet and the in port of the detector.
3
Set up the instrument with the following test conditions.
When
If you want to checkout the detector
Tools required Description
LC system with G7162A RID
Parts required # p/n Description
1 G1362-87301 Restriction capillary
Table 10 Chromatographic Conditions
Mobile phases LC grade water
Column Restriction capillary 2.7 m x 1.7 mm i.d.
Flow rate 1.0 mL/min
Compressibility 46
Stroke 20 µL
Stop time 20 min
Column compartment temperature 35 °C
Optical unit temperature 35 °C
Polarity Positive
Peak width (response time) 0.2 min (4 s, standard)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals