Agilent InfinityLab LC Series RID User Manual 119
8 Test Functions
ASTM Drift and Noise Test
ASTM Drift and Noise Test
Setting the Test Conditions
1
Turn
ON
the detector.
You are now ready to change the settings of your detector.
2
Connect the restriction capillary directly between the column compartment
heat exchanger outlet and the in port of the detector.
3
Set up the instrument with the following test conditions.
When
If you want to checkout the detector
Tools required Description
LC system with G7162A RID
Parts required # p/n Description
1 G1362-87301 Restriction capillary
Table 10 Chromatographic Conditions
Mobile phases LC grade water
Column Restriction capillary 2.7 m x 1.7 mm i.d.
Flow rate 1.0 mL/min
Compressibility 46
Stroke 20 µL
Stop time 20 min
Column compartment temperature 35 °C
Optical unit temperature 35 °C
Polarity Positive
Peak width (response time) 0.2 min (4 s, standard)