Agilent InfinityLab LC Series RID User Manual 37
2 Site Requirements and Specifications
Physical Specifications
Physical Specifications
Table 1 Physical Specifications
Type Specification Comments
Weight 14.7 kg (32.4 lbs)
Dimensions
(height × width × depth)
180 x 396 x 436 mm
(7.1 x 15.6 x 17.2 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 80 VA / 70 W
Ambient operating tempera-
ture
4–55 °C (39–131 °F)
Ambient non-operating tem-
perature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Safety standards: IEC, EN,
CSA, UL
Installation category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11