Agilent InfinityLab LC Series RID User Manual 88
7Error Information
General Error Messages
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 4 of the remote connector generates the error
message.
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one minute
of starting the analysis. If a not-ready condition is still present on the remote line
after one minute the error message is generated.
Probable cause Suggested actions
1
Leak detected in an external instrument with
a remote connection to the system.
Fix the leak in the external instrument before
restarting the module.
2
Shut-down in an external instrument with a
remote connection to the system.
Check external instruments for a shut-down con-
dition.
3
The degasser failed to generate sufficient
vacuum for solvent degassing.
Check the vacuum degasser for an error condi-
tion. Refer to the Service Manual for the degasser
or the pump that has the degasser built-in.
Probable cause Suggested actions
1
Not-ready condition in one of the instru-
ments connected to the remote line.
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up cor-
rectly for analysis.
2
Defective remote cable.
Exchange the remote cable.
3
Defective components in the instrument
showing the not-ready condition.
Check the instrument for defects (refer to the
instrument’s documentation).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals