EasyManuals Logo

Agilent Technologies 1260 Infinity II Refractive Index Detector User Manual

Agilent Technologies 1260 Infinity II Refractive Index Detector
217 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Agilent InfinityLab LC Series RID User Manual 95
7Error Information
Refractive Index Detector Specific Error Messages
Heater Fuse
Error ID: 2682
The electronic fuse of the heater has been activated.
Wrong Temperature Profile
Error ID: 2683
After turning ON the optical unit heat control, the temperature does not increase
at a sufficiently fast rate to reach the set point.
Probable cause Suggested actions
1
Short in heater circuit.
Powercycle the module.
2
Defective main board.
Please contact your Agilent service representa-
tive.
3
Defective heater.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Defective main board.
Please contact your Agilent service representa-
tive.
2
Defective heater.
Please contact your Agilent service representa-
tive.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals