Agilent InfinityLab LC Series RID User Manual 38
2 Site Requirements and Specifications
Performance Specifications
Performance Specifications
Table 2 Agilent 1260 Infinity II Refractive Index Detector (G7162A) Performance Specifications
Type Specification
Detection type Refractive Index
Refractive index range 1.00 – 1.75 RIU, calibrated
Measurement range
±600·10
-6
RIU
Optical zeroing Digitally via software, manually via set screw
Optics temperature control 5 °C above ambient to 55 °C
Sample cell Volume: 8 µL
Maximum pressure: 5 bar (0.5 MPa)
Maximum flow rate: 5 mL/min
Valves Automatic purge and automatic solvent recycle
Volumes Inlet port to sample cell 62 µL,
inlet port to outlet port 590 µL
Liquid contact materials 316 stainless steel, PTFE
and quartz glass
pH range 2.3 – 9.5
Performance specifications
Short term noise: <±1.25·10
-9
RIU
Drift: <200·10
-9
RIU/hr
Time programmable param-
eters
Polarity, peak width
Maximum data rate 74 Hz
Detector zero Automatic zero before analysis
Instrument Control Lab Advisor B.02.07 or above
LC and CE Drivers A.02.12 or above
Local control Agilent Instant Pilot (G4208A)
Analog outputs Recorder/integrator: 100 mV or 1 V, output range selectable, one out-
put
Communication LAN, Controller Area Network (CAN), Extended Remote Interface
(ERI): ready, start, stop and shut-down signals