EasyManuals Logo

Agilent Technologies 1260 Infinity II Refractive Index Detector User Manual

Agilent Technologies 1260 Infinity II Refractive Index Detector
217 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
Agilent InfinityLab LC Series RID User Manual 98
7Error Information
Refractive Index Detector Specific Error Messages
Purge Valve Not Connected
Error ID: 2688
When activated no response was received from the purge valve.
Recycle Valve Not Connected
Error ID: 2689
When activated no response was received from the recycle valve.
Probable cause Suggested actions
1
Purge valve disconnected.
Please contact your Agilent service representa-
tive.
2
Defective purge valve.
Please contact your Agilent service representa-
tive.
3
Defective main board.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Recycle valve disconnected.
Please contact your Agilent service representa-
tive.
2
Defective recycle valve.
Please contact your Agilent service representa-
tive.
3
Defective main board.
Please contact your Agilent service representa-
tive.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals