Agilent InfinityLab LC Series RID User Manual 109
8 Test Functions
Refractive Index Calibration
Preparing the Calibration
1
Preparation of the sucrose calibration solution.
a
To prepare 25 ml of the calibration solution 87.5 mg of the Sucrose
sample is required.
b
Add the weighed amount of sample into a suitable volumetric flask.
c
Dispense 10 ml of LC grade water into the flask and shake or stir to
dissolve.
d
Dilute the contents of the flask to volume with LC grade water.
Wait five minutes and shake again. The solution is now ready for use.
2
Preparing the pump.
a
Fill a suitable solvent bottle with LC grade water.
b
Connect this bottle to Channel A of the pump, A1 if a binary pump.
3
Flush the degasser and pump.
When
Recommended after exchange of the optical unit or RIM board.
Tools required Description
Laboratory balance
Parts required # p/n Description
1 DAB/Ph Eur/BP/JP/NF/USP Grade Sucrose
1 9301-1446 Syringe
1 9301-0407 Needle
1 5190-5111 Syringe filter, 0.45 µm, 100/pk
1 0100-1516 Fingertight fitting PEEK, 2/pk

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals