Agilent InfinityLab LC Series RID User Manual 177
12 Hardware Information
Setting the 6-bit Configuration Switch
Table 15 6-bit Configuration Switch
Mode Function/Setting
Switch 1 Switch 2 Switch 3 Switch 4 Switch 5 Switch 6
COM
1
1
When selecting mode COM, settings are stored to non-volatile memory. When selecting mode TEST, COM settings are taken from
non-volatile memory.
0
n.a.
2
2
not assigned - Always keep these switches on position ‘0’ (off)
n.a. LAN Init Mode n.a.
Use Default IP Address
3
3
Default IP Address is 192.168.254.11
00000
Use Stored IP Address00010
Use DHCP to request IP Address
4
4
Host Name will be the MAC address.
00100
Test 1 System n.a. n.a. n.a. ColdStart
Boot Main System / Keep Data00000
Boot Resident System / Keep Data10000
Boot Main System / Revert to
Default Data
00001
Boot Resident System / Revert to
Default Data
10001

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals