Agilent InfinityLab LC Series RID User Manual 54
3Using the Module
Detector Control Settings
Purge Reference Cell: This parameter is used to exchange the content of the
reference cell in the case of solvent change or reference cell contamination.
Enter a time interval (minutes) to purge the reference cell. This will be started
immediately if you click OK on this window. Allow additional time for baseline
stabilization after purging.
•Recycling Valve: Select the On option to switch the recycling of the eluent on.
The Off option diverts the flow of the RID to the waste bottle.
•Automatic Turn On: This function allows you to turn on the optical unit heater
at a specified date and time. It requires that the At Power On function is turned
off. Select Turn Heater on at: to activate the date and time fields, and enter the
date and time in the appropriate fields in the specified format.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity II Refractive Index Detector and is the answer not in the manual?

Agilent Technologies 1260 Infinity II Refractive Index Detector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity II Refractive Index Detector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals