Chapter 3 Features and Functions
Output Configuration
3
•For a triggered burst, the Sync signal is a TTL “high” when the burst
begins
. The Sync signal
is a TTL “low” at the end of the specified
number of cycles (may not be
the zero-crossing point if the waveform
has an associated start phase). For an
infinite count burst
, the Sync
signal is the same as for a continuous waveform.
For an externally-gated burst, the Sync signal follows the external
gate signal. However, the signal will not go to a TTL “low” until the
end of the last cycle (may not be the zero-crossing point if the
waveform has an associated start phase).
Front-Panel Operation:Press
and press the Sync softkey
again to toggle between Off and On. You can also press Sync Setup to
configure Sync.
Remote Interface Operation:
OUTPut:SYNC {OFF|ON}
OUTPUT[1|2]:SYNC:MODE {CARRier|NORMal|MARKer}
OUTPUT[1|2]:SYNC:POLARITY {NORMAL|INVerted}
OUTPUT:SYNC:SOURCE {CH1|CH2}

Table of Contents

Other manuals for Agilent Technologies 33500 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33500 Series and is the answer not in the manual?

Agilent Technologies 33500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model33500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals