138
Chapter 3 Features and Functions
Pulse Width Modulation (PWM)
3
Pulse Waveform
Pulse is the only waveform shape supported for PWM.
Front-Panel Operation:Press
, then
Pulse
.
Remote Interface Operation:
FUNCtion PULSe
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Pulse Period
The range for the pulse period is 33 ns to 1,000,000 s. The default is 100
s.
Front-Panel Operation: After selecting the pulse function, press
and the Frequency/Period softkey to toggle to the Period
setting. Then use the knob or numeric keypad to enter the desired
pulse period. If you use the numeric keypad, press a softkey to specify
the units.
Remote Interface Operation:
FUNCtion:PULSe:PERiod {<seconds>|MINimum|MAXimum}