EasyManua.ls Logo

Agilent Technologies 33500 Series - Page 139

Agilent Technologies 33500 Series
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
138
Chapter 3 Features and Functions
Pulse Width Modulation (PWM)
3
Pulse Waveform
Pulse is the only waveform shape supported for PWM.
Front-Panel Operation:Press
, then
Pulse
.
Remote Interface Operation:
FUNCtion PULSe
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Pulse Period
The range for the pulse period is 33 ns to 1,000,000 s. The default is 100
s.
Front-Panel Operation: After selecting the pulse function, press
and the Frequency/Period softkey to toggle to the Period
setting. Then use the knob or numeric keypad to enter the desired
pulse period. If you use the numeric keypad, press a softkey to specify
the units.
Remote Interface Operation:
FUNCtion:PULSe:PERiod {<seconds>|MINimum|MAXimum}

Table of Contents

Other manuals for Agilent Technologies 33500 Series

Related product manuals