154
Chapter 3 Features and Functions
Burst Mode
3
When the gated mode is selected, the burst count, burst period,
and trigger source are ignored (these parameters are used for the
triggered burst mode only). If a manual trigger is received, it is
ignored and no error will be generated.
When the gated mode is selected, you can also select the polarity of
the signal on the rear-panel Ext Trig connector.
Front-Panel Operation: After enabling bursts, press the N Cycle
(triggered) or Gated softkey.
Remote Interface Operation:
[SOURce[1|2]:]BURSt:MODE {TRIGgered|GATed}
Use the following command to select the polarity of the external gate
signal on the
Ext Trig
connector. The default is NORM (true-high
logic).
[SOURce[1|2]:]BURSt:GATE:POLarity {NORMal|INVerted}