Chapter 3 Features and Functions
Burst Mode
3
Burst Period
The burst period defines time from the start of one burst to the start of
the next burst. Used in the internal triggered burst mode only.
Keep in mind that burst period is different than the “waveform frequency”
which specifies the frequency of the bursted signal.
•Burst period:1s to 8000 seconds. The default is 10 ms.
The burst period setting is used only when Immediate triggering is
enabled. The burst period is ignored when manual or external
triggering is enabled (or when the gated burst mode is selected).
•It is not possible to specify a burst period which is too short for the
waveform generator to output with the specified burst count and
frequency (see below). If the burst period is too short, the waveform
generator will automatically adjust it as needed to continuously
re-trigger the burst.
Front-Panel Operation
: To set the burst period, press
and
then
the
Burst Period
softkey. Then use the knob or numeric keypad to
enter the period.
Remote Interface Operation:
[SOURce[1|2]:]BURSt:INTernal:PERiod
{<seconds>|MINimum|MAXimum}

Table of Contents

Other manuals for Agilent Technologies 33500 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33500 Series and is the answer not in the manual?

Agilent Technologies 33500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model33500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals