163
Chapter 3 Features and Functions
Triggering
3
External Triggering In the external trigger mode, the waveform
generator will accept a hardware trigger applied to the rear-panel
Trig In connector. The waveform generator initiates one sweep or
outputs one burst each time Ext Trig receives a TTL pulse with the
specified edge.
See also “Trigger Input Signal,” on the following page.
Front-Panel Operation: The external trigger mode is like the manual
trigger mode except that you apply the trigger to the
Ext Trig
connector
. To select the external source, press
and then
Trg Src and Ext.
To specify whether the waveform generator triggers on the rising or
falling edge, press the Trigger Setup softkey and select the desired
edge direction by pressing Slope.
Remote Interface Operation:
TRIGger:SOURce EXTernal
Use the following command to specify whether the waveform
generator triggers on the rising or falling edge.
TRIGger:SLOPe {POSitive|NEGative}

Table of Contents

Other manuals for Agilent Technologies 33500 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33500 Series and is the answer not in the manual?

Agilent Technologies 33500 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model33500 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals