Chapter 2 Front-Panel Menu Operation
To Output an FSK Waveform
2
3 Set the hop frequency.
Press the Hop Freq softkey and then set the value to 500 Hz using the
numeric keypad or the knob and cursor keys. If you use the numeric
keypad, be sure to finish the entry by pressing the Hz softkey.
4 Set the FSK shift rate.
Press the FSK Rate softkey and then set the value to 100 Hz using the
numeric keypad or the knob and cursor keys.
At this point, the waveform generator outputs an FSK waveform.