6 General Maintenance
To Clean the EI HES
158 5977C GC/MSD Operating Manual
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
3 If the contamination is serious, such as an oil backflow into the analyzer,
seriously consider replacing the contaminated parts
CAUTION
If insulators are dirty, clean them with a cotton swab dampened with
reagent-grade methanol. If that does not clean the insulators, replace them. Do
not abrasively or ultrasonically clean the insulators.
Figure 49 EI HES parts to be cleaned
Source body
Repeller
Entrance
lens (1)
Ion focus
lens (2)
Post extractor
lens 2 (3)
Post extractor
lens 1 (4)
Extractor
lens (5)
CAUTION
The filaments, source heater assembly, insulators, source mounting plate, and
filament block cannot be cleaned ultrasonically. Replace these components if
major contamination occurs.
CAUTION
Do not use the abrasive slurry on the source mount bushings.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Series and is the answer not in the manual?

Agilent Technologies 5977C Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals