6 General Maintenance
To Clean the EI HES
5977C GC/MSD Operating Manual 159
4 Abrasively clean the surfaces that contact the sample or ion beam.
Use an abrasive slurry of alumina powder and reagent-grade methanol on a
cotton swab. Use enough force to remove all discolorations. Polishing the
parts is not necessary; small scratches will not harm performance. Also,
abrasively clean the discolorations where electrons from the filaments
enter the source body.
5 Rinse away all abrasive residue with reagent-grade methanol.
Ensure all abrasive residue is rinsed away before ultrasonic cleaning. If the
methanol becomes cloudy or contains visible particles, rinse again.
6 Separate the parts that were abrasively cleaned from the parts that were not
abrasively cleaned.
7 Ultrasonically clean the parts (each group separately) for 15 minutes in each
of the following solvents:
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Methanol (reagent-grade)
8 Place the parts in a clean beaker. Loosely cover the beaker with clean
aluminum foil (dull side down).
9 Dry the cleaned parts in an oven at 100 °C for 5–6 minutes.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
WARNING
All of these solvents are hazardous. Work in a fume hood, and take all
appropriate precautions.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977C Series and is the answer not in the manual?

Agilent Technologies 5977C Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977C Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals