EasyManua.ls Logo

Agilent Technologies 6890 Series - Page 24

Agilent Technologies 6890 Series
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Site Preparation
Cryogenic cooling requirements
20
These contaminants could clog the expansion orifice or affect the proper
operation of the GC.
Additional requirements for the liquid CO
2
system include:
The tank must have an internal dip tube or eductor tube to deliver liquid CO
2
instead of gas (see Figure 5).
The liquid CO
2
must be provided to the GC at a pressure of 700 to 1,000 psi
at a temperature of 25°C.
Use 1/8-inch diameter heavy-wall stainless steel tubing for supply tubing. The
tubing should be between 5 to 50 feet long.
Coil and fasten the ends of the tubing to prevent it from “whipping” if it
breaks.
Do not install a pressure regulator on the CO
2
tank, as vaporization and
cooling would occur in the regulator instead of the oven.
Do not use a padded tank (one to which another gas is added to increase the
pressure).
Figure 5. Correct and incorrect liquid CO
2
tank configuration
WARNING Do not use copper tubing or thin-wall stainless steel tubing with liquid CO
2
. Both
harden at stress points and may explode.
Dip tube
Correct configuration Incorrect configuration

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Related product manuals