Site Preparation
Gas requirements
12
Table 4. Gas Recommendations for Packed Columns
Gases for capillary columns
When used with capillary columns, GC detectors require a separate makeup gas
for optimum sensitivity. For each detector and carrier gas, there is a preferred
choice for makeup gas. Table 5 lists gas recommendations for capillary columns.
Detector Carrier gas Comments
Detector, anode purge,
or reference gas
Electron
Capture
Nitrogen Maximum sensitivity Nitrogen
Argon/Methane Maximum dynamic range Argon/Methane
Flame
Ionization
Nitrogen Maximum sensitivity Hydrogen and
air for detector
Helium Acceptable alternative
Flame
Photometric
Hydrogen H ydr oge n a nd a ir fo r d e te ct o r
Helium
Nitrogen
Argon
Nitrogen-
Phosphorus
Helium Optimum performance Hydrogen and
air for detector
Nitrogen Acceptable alternative
Thermal
Conductivity
Helium General use Reference must be same as
carrier
Hydrogen Maximum sensitivity (Note A)
Nitrogen Hydrogen detection (Note B)
Argon Maximum hydrogen sensitivity
(Note B)
Note A: Slightly greater sensitivity than helium. Incompatible with some compounds.
Note B: For analysis of hydrogen or helium. Greatly reduces sensitivity for other compounds.

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals