EasyManuals Logo
Home>Agilent Technologies>Portable Generator>83640B

Agilent Technologies 83640B Service Guide

Agilent Technologies 83640B
603 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #193 background imageLoading...
Page #193 background image
3. On
the oscilloscop
e, set:
Channel
1:
Display On
Volts/Division
500
mV
Input
Coupling dc
Input
Impedance
50

Timebase:
Time/Division
10
ns
Delay
Reference
center
Sweep
auto
Trigger:
Trigger
Mode
Edge
Trig
Src
Chan
1
Trigger
Level
1.2V
Slope
Pos
Display:
Display Mode
Repetitive
Averaging
Off
Display
Time
0.2
s
4.
Adjust
the
oscilloscop
e
oset
to
fully
displa
y
the
signal.
5.
Use
the oscilloscop
e
to
measure
the
width
of
the
pulse.
Record
the
v
alue
measured
on
the
test
record.
6.
On
the
syn
thesizer,
c
hange
the
pulse
width
to
50
ns.
Press:
N
N
N
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
N
N
N
N
N
N
N
Internal
Pulse
Width
4
50
5
4
ns
5
7.
Use
the
oscilloscop
e
to
measure
the
width
of
the
pulse. Record
the v
alue
measured
on
the
test
record.
Related
Adjustments
None
In
Case
of
Difficulty
Refer
to
the
A
gilent
T
e
chnologies
8360
B-Series
Swept
Signal
Gener
ator/
8360
L-Series
Swept CW Gener
ator T
roubleshooting Guide
.
2-106 Performance Tests
Internal Pulse Accuracy
Agilent 8360

Table of Contents

Other manuals for Agilent Technologies 83640B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83640B and is the answer not in the manual?

Agilent Technologies 83640B Specifications

General IconGeneral
BrandAgilent Technologies
Model83640B
CategoryPortable Generator
LanguageEnglish

Related product manuals