Table
2-34. T
est Record
for 83623L
(4 of
5)
Mo
del
Rep
ort No.
Date
T
est
Minim
um
Maxim
um
Measuremen
t
No. T
est
Description
Sp
ec.
Results Sp
ec.
Uncertain
t
y
7. External
Leveling
7.
Minim
um P
ow
er
0
0.2
mV
6
3
V
10.
Maximum
Po
wer
0
0.5 V
6
20
V
8. Spurious
Signals: Harmonic
Harmonics:
8.
<
50
MHz
0
20
dBc
6
1.75
dB
8.
50
MHz
to
<
2.0
GHz
0
25
dBc
6
1.75
dB
12.
2.0
to
20
GHz
0
45
dBc
6
1.75
dB
Subharmonics
12.
7
to
20
GHz
0
50
dBc
6
1.75
dB
9. Spurious
Signals:
Non-harmonic
11. Spur
Frequency:
0
60
dBc
6
1
dB
0
60
dBc
6
1
dB
0
60
dBc
6
1
dB
0
60
dBc
6
1
dB
125
kHz
0
60 dBc
6
1dB
500 kHz
0
60
dBc
6
1
dB
16.
Spur
F
requency:
100
MHz
Fixed
0
60
dBc
6
1
dB
100
MHz
Oset
0
60
dBc
6
1
dB
LO
F
eedthrough
0
60
dBc
6
1
dB
20.
Lo
w
Band
Mixer
Spur
0
60
dBc
6
1.85
dB
2-152 Performance Tests