108 Chapter2
Making Adjustments
16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
3. Press INSTRUMENT PRESET on the synthesized sweeper. Set the
synthesized sweeper controls as follows:
CW ................................................................ 300MHz
FREQ STEP ................................................... 72MHz
POWER LEVEL ............................................. −3dBm
4. To enter band 0, press the following analyzer keys.
PRESET
CAL, More 1 of 4
CORRECT ON OFF (OFF)
FREQUENCY, BAND LOCK, 0 - 2.9 Gz BAND 0
(Skip these key presses for the 8594E, 8594L and 8594Q
analyzers.)
5. Set the analyzer controls by pressing the following keys.
FREQUENCY, 300, MHz
CF STEP AUTO MAN (MAN), 72, MHz
SPAN, 10, MHz
AMPLITUDE, 0, dBm
SCALE LOG LIN (LOG), 10, dB
BW, 1, MHz
PEAK SEARCH
MKR FCTN, MKR TRACK ON OFF (ON)
6. Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK
amplitude reading of −9 dBm ±0.1 dB on the analyzer.
7. Press RATIO on the measuring receiver.
Band 0
8. Set the synthesized sweeper CW to 12 MHz.
9. Press the following analyzer keys.
FREQUENCY, 12, MHz
10.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK
amplitude reading of −9 dBm ±0.1 dB on the analyzer.
11.Record the power ratio displayed on the measuring receiver in
Column 2 of Table 2-8.
12.On the synthesized sweeper, press CW and ⇑ (step-up key).