Chapter 2 109
Making Adjustments
16. Frequency Response of the 8592L/94L, 8593E/94E/95E/96E, and 8594Q
13.On the analyzer, press the following keys.
FREQUENCY, ⇑
14.Step through the remaining frequencies listed in Table 2-8. At each
new frequency, repeat step 10 to step 13 and enter the appropriate
power sensor cal factor into the measuring receiver as listed in
Column 3 of Table 2-8.
Adjustment for the 8594E and 8594L spectrum analyzer and 8594Q
QAM analyzer is complete. Continue with “Entering Flatness
Correction Data” at the end of this adjustment.
Continue with the next step for all other analyzers.
Band 1
15.Replace the 1 MHz to 2.9 GHz power sensor with the 50 MHz to
26.5 GHz power sensor.
16.Calibrate the measuring receiver and 50 MHz to 26.5 GHz power
sensor combination in log mode.
17.Set the synthesized sweeper CW to 300 MHz.
18.Press the following analyzer keys.
FREQUENCY, 300, MHz
19.Adjust the synthesized sweeper POWER LEVEL for a MKR-TRK
amplitude reading of −9 dBm ±0.1 dB on the analyzer.
20.Press RATIO on the measuring receiver.
21.To enter band 1, press the following analyzer keys.
FREQUENCY, BAND LOCK, 2.75 - 6.4 BAND 1
22.Set the analyzer controls by pressing the following keys.
FREQUENCY, 2.75, GHz
CF STEP AUTO MAN (MAN), 234.9, MHz
SPAN, 10, MHz
BW, 1, MHz
23.Set the synthesized sweeper CW to 2.75 GHz and FREQ STEP to
234.3 MHz.
24.Press the following analyzer keys.
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
AMPLITUDE, PRESEL PEAK