1-230 Chapter1
Alphabetical Command Reference
SET
SET
Syntax
SET<BIT|Z><num>; or SET<BIT|Z>?;
SET<DATE|TIME><$>;
SETF;
SET<RREFL|RRTHRU>; or SET<RREFL|RRTHRU>?;
Description
Front Panel Equivalents
Command Description Range Query Response
SETBIT Sets the specified bit on the GPIO. integers 0–7
<num><
L
F
>
SETDATE Sets the date in the following format: DD
MMM YYYY, where DD is the day and
must be 2 digits, MMM is the month and
must be three alpha characters (JAN,
FEB, MAR, APR, MAY, JUN, JUL, AUG,
SEP, OCT, NOV, DEC), and YYYY is the
year and must be 4 digits.
See “Description. N/A
SETF Sets the frequency for low pass transform,
Option 010.
N/A N/A
SETRTHRU
1
1. ES models only.
Sets the reference thru for a TRL cal. N/A
<0|1><
L
F
>
SETRREFL
1
Sets the reference reflect for a TRL cal. N/A
<0|1><
L
F
>
SETTIME Sets the time in the following format:
HH:MM:SS, where HH is the hour, MM is
minutes, SS is seconds, and each must be
2 digits.
See “Description. N/A
SETZ Sets the characteristic impedance of the
measurement system.
0.1 to 500
<num><
L
F
>
Command Hardkey Softkey
SETBIT
SETDATE
Seq
SET BIT
System
SET DAY
SET MONTH
SET YEAR

Table of Contents

Other manuals for Agilent Technologies 8719ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8719ET and is the answer not in the manual?

Agilent Technologies 8719ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8719ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals