EasyManuals Logo

Agilent Technologies 8719ET Programmer's Guide

Agilent Technologies 8719ET
541 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #254 background imageLoading...
Page #254 background image
1-240 Chapter1
Alphabetical Command Reference
SPEC
SPEC
Syntax
SPEC<$><num,num,…,num>;
Description
The following commands initiate the SPECIFY CLASS part of modifying a cal kit. With
each command, send the analyzer a series of standard numbers to be included in the class.
When the class is full, send the CLAD command to terminate the specification.
Command Description Range for
<num,num,...>
Query Response
SPECFWDM Forward match standard numbers
1–8
N/A
SPECFWDT Forward transmission standard numbers
1–8
N/A
SPECRESP Response standard numbers
1–8
N/A
SPECRESI For Resp & Isol, specifies the response
standards
standard numbers
1–8
N/A
SPECREVM Reverse match standard numbers
1–8
N/A
SPECREVT Reverse transmission standard numbers
1–8
N/A
SPECS11A S11A standard numbers
1–8
N/A
SPECS11B S11B standard numbers
1–8
N/A
SPECS11C S11C standard numbers
1–8
N/A
SPECS22A S22A standard numbers
1–8
N/A
SPECS22B S22B standard numbers
1–8
N/A
SPECS22C S22C standard numbers
1–8
N/A
SPECTRLL TRL Line or Match standard numbers
1–8
N/A
SPECTRLT TRL Thru standard numbers
1–8
N/A

Table of Contents

Other manuals for Agilent Technologies 8719ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8719ET and is the answer not in the manual?

Agilent Technologies 8719ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8719ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals