EasyManuals Logo

Agilent Technologies 8719ET User Manual

Agilent Technologies 8719ET
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #398 background imageLoading...
Page #398 background image
7-28
Operating Concepts
Analyzer Display Formats
Figure 7-11 Linear Magnitude Format
SWR Format
The softkey reformats a reflection measurement into its equivalent SWR (standing
wave ratio) value. See Figure 7-12. SWR is equivalent to (1 + ρ)/(1 ρ), where ρ is the
reflection coefficient. Note that the results are valid only for reflection measurements. If
the SWR format is used for measurements of S
21
or S
12
the results are not valid.
Figure 7-12 Typical SWR Display
SWR

Table of Contents

Other manuals for Agilent Technologies 8719ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8719ET and is the answer not in the manual?

Agilent Technologies 8719ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8719ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals