EasyManuals Logo

Agilent Technologies 8719ET User Manual

Agilent Technologies 8719ET
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #426 background imageLoading...
Page #426 background image
7-56
Operating Concepts
Calibration Routines
E-CAL
The E-Cal calibration menu is activated by pressing in the calibration
menu.
The E-Cal (Electronic Calibration) system determines systemic errors of the analyzer
through a one-time connection of an E-Cal module to the network analyzer ports. The
random error of connector repeatability is reduced substantially through a one-time
connection when compared to frequent connections and disconnections of the conventional
short/open/load methods.
E-CAL MENU

Table of Contents

Other manuals for Agilent Technologies 8719ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8719ET and is the answer not in the manual?

Agilent Technologies 8719ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8719ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals