Chapter 2 135
Front-Panel Key Reference
Meas Setup
Avg Mode 
Exp Repeat
Allows you to select the type of termination control used for the 
averaging function. This determines the averaging action after the 
specified number of measurements (average count) is reached. 
Avg Mode has no effect on measurements that are not averaged.
Exp  Each successive data acquisition after the average 
count is reached is exponentially weighted and 
combined with the existing average. Exponential 
averaging weights new data more than old data, which 
facilitates tracking of slow-changing signals.
Repeat  After reaching the average count, all previous result 
data is cleared and the average count is set back to 1. 
Key Access: 
Meas Setup
2nd Car Offs
Allows you to change the frequency difference from the main carrier to 
the second carrier, measured in MHz. The center frequency of the main 
carrier is the analyzer center frequency less half of the second carrier 
offset in MHz. The second carrier center frequency is the main carrier 
center frequency plus the second carrier offset value in MHz. The 
available offset values are –15, –10, –5, 5, 10, and 15 MHz. The 
negative values set the offset below the main carrier frequency, and the 
positive values set the offset above the main carrier frequency.
Key Access: 
Meas Setup
Optimize
Ref Level
Sets the reference level and attenuator to optimum values based on the 
signal present at the input. 
Key Access: 
Meas Setup
Offset/Limits
Displays menu keys that allow you to configure the offsets for the 
multi-carrier power measurement.
Key Access: 
Meas Setup, More
Offset
A B C D
  Selects the offset the menu keys will affect. Press Offset 
until the letter of the desired offset (A, B, C, or D) is 
underlined.
Upper Offset
Limit
  Sets the limit for the upper segment of the specified 
offset pair. 
Lower Offset
Limit
  Sets the limit for the lower segment of the specified 
offset pair.
Key Access: 
Meas Setup, More, Ofs & Limits