Chapter 2 147
Front-Panel Key Reference
Meas Setup
Spectrum Emission Mask Meas Setup Menu Keys
To access the following keys for setting up a spectrum emission mask 
measurement, press 
MEASURE, Spectrum Emission Mask, and then the 
front-panel 
Meas Setup key.
Avg Number 
On Off
Press Avg Number (On) to specify the number of averages used when 
calculating the measurement result. The average will be displayed at 
the end of each sweep. 
Avg Number (Off) disables measurement 
averaging.
Key Access: 
Meas Setup
Meas Type
Allows you to specify the reference for the spectrum emission mask 
measurement. 
Total Pwr Ref  Press Total Pwr Ref to specify the reference for the 
measurement to the total power.
PSD Ref  Press PSD Ref to specify the reference for the 
measurement to power spectral density.
Key Access: 
Meas Setup, Meas Type
Ref Channel
Chan Integ BW
  Allows you to specify the channel integration 
bandwidth used in calculating the power in the main 
channel.
Chan Span  Allows you to specify the channel span used in 
calculating the power in the main channel.
Sweep Time
Auto Man
   Allows you to specify the sweep time used in calculating 
the power in the main channel.
Res BW
Auto Man
  Allows you to specify the resolution bandwidth used in 
calculating the power in the main channel.
Total Pwr Ref
Auto Man
  This parameter allows you to select whether to 
calculate the carrier power based on the signal at the 
analyzer input, or use a fixed value as reference. 
Enables the use of the Agilent Dynamic Range 
Extender. The Agilent Dynamic Range Extender will 
effectively “remove” the carrier from the signal coming 
into the analyzer. The fixed value can be entered 
manually.
Key Access: 
Meas Setup, Ref Channel