Chapter 2 139
Front-Panel Key Reference
Meas Setup
Burst Power Meas Setup Menu Keys
To access the following keys for setting up a burst power measurement, 
press 
MEASURE, Burst Power and then the front-panel Meas Setup key.
Key Access: 
Meas Setup
Avg Number 
On Off
Press Avg Number (On) to specify the number of averages used when 
calculating the measurement result. The average will be displayed at 
the end of each sweep. 
Avg Number (Off) disables measurement 
averaging.
Key Access: 
Meas Setup
Avg Mode 
Exp Repeat
Allows you to select the type of termination control used for the 
averaging function. This determines the averaging action after the 
specified number of measurements (average count) is reached. 
Avg Mode has no effect on measurements that are not averaged.
Exp  Each successive data acquisition after the average 
count is reached is exponentially weighted and 
combined with the existing average. Exponential 
averaging weights new data more than old data, which 
facilitates tracking of slow-changing signals.
Repeat  After reaching the average count, all previous result 
data is cleared and the average count is set back to 1. 
Key Access: 
Meas Setup
Avg Type 
Video Power
Allows you to specify the type of result averaging to be performed.
Video  Selects averaging that sums the trace data and divides 
by the number of data .
Power  Selects averaging that converts trace data from dBm to 
power units, then averages the power trace data. This 
selection requires more time to perform.
Key Access: 
Meas Setup
Threshold Lvl
Abs Rel
Allows you to set the level above which the mean carrier power 
calculation is based upon. The threshold level may be described in dB 
(relative to the carrier) or dBm (absolute). 
Key Access: 