194 Chapter 2 
Front-Panel Key Reference
Sweep
Sweep 
Single Cont
Switches the analyzer between the continuous-sweep mode and the 
single-sweep mode. Pressing 
Sweep (Single) puts the analyzer in 
single-sweep mode. Press 
Single on the front panel, to enable a sweep 
when in single-sweep mode. When 
Sweep (Cont) is selected, one sweep 
follows another as soon as it is triggered. Pressing 
Preset (When Preset 
Type
 is set to Factory) or turning the power on, sets the sweep mode to 
continuous sweep.
Key Access: 
Sweep
Auto Swp 
Coupling 
SR SA
Selects stimulus-response (SR) or spectrum-analyzer (SA) (default) 
auto-coupled sweep times. In stimulus-response mode, auto-coupled 
sweep times are usually much faster for swept-response measurements. 
Stimulus-response auto-coupled sweep times are typically valid in 
stimulus-response measurements when the system’s frequency span is 
less than 20 times the bandwidth of the device under test. 
Key Access: 
Sweep
Gate
Requires Option 1D6 (Time Gate). Accesses the following menu keys 
that allow you to setup various gate parameters: 
Key Access: 
Sweep
Gate On Off
  Requires Option 1D6 (Time Gate). Turns the gate 
function on and off. When set to 
Gate (On), the video 
signal that is digitized is controlled by the gate 
circuitry. The gate circuitry switches between two 
states. When the gate is “open”, the normal video signal 
of the analyzer is passed through the video filters to the 
peak detectors and digitizer of the analyzer. When the 
gate is “closed”, the video filters, peak detectors, and 
digitizer are given a signal at the bottom of the display.
The gate function requires that a gate trigger signal be 
connected to the 
GATE TRIG/EXT TRIG IN (TTL) input on 
the rear panel. When the gate function is on, the state 
of the gate appears at the 
GATE/HI SWP OUT (TTL) rear 
panel connector. A TTL high output indicates that the 
gate is open. The gate out signal is only valid while the 
analyzer is sweeping. Between sweeps, the gate out 
signal is invalid. You can adjust the gate delay and gate 
length using an oscilloscope to view the gate out signal. 
The analyzer sweep time should be set to a large value 
(50 sec) during the adjustment so that the gate signal is 
valid most of the time. When the adjustment is done 
you can set the analyzer back to the desired sweep 
time, or set it back to 
Auto Couple.
Key Access: 
Sweep, Gate