210 Chapter 2
Front-Panel Key Reference
System (Local)
Licensing Accesses the security system to enable licensing for individual options.
For information on how to use this system, see the installation
information for individual options in the measurement personalities’
user’s guides.
Table 2-9 System Default Settings
Feature Default Setting
Color Printing On
Custom Printer, Color Capability No
Custom Printer, Language PCL3
Date Format MDY
Display Viewing Angle 4
External Amplifier Gain 0 dBm
GPIB Address 18
Input Impedance
50 (Ohm)
1
Instrument Color Palette Default palette
Manual Tracking Adjustment 2048
Page Size Letter
Power On Last/Preset Preset
Preset Factory/User Mode
Print Format Print screen mode
Print Orientation Portrait
Printer Selection Auto
Prints/ Page 1
RS-232 Baud Rate 9600
RS-232 DTR Control On (hardware pacing)
RS-232 Receive Pacing None (data pacing)
RS-232 RTS Control IBFULL (hardware pacing)
RS-232 Transmit Pacing None (data pacing)
Speaker Volume 0
Time Date Display On
Verbose SCPI Off
1. This value may be 75 ohms, depending upon the analyzer hardware config-
uration.