EasyManuals Logo

Agilent Technologies E4402B User Manual

Agilent Technologies E4402B
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #210 background imageLoading...
Page #210 background image
210 Chapter 2
Front-Panel Key Reference
System (Local)
Licensing Accesses the security system to enable licensing for individual options.
For information on how to use this system, see the installation
information for individual options in the measurement personalities’
users guides.
Table 2-9 System Default Settings
Feature Default Setting
Color Printing On
Custom Printer, Color Capability No
Custom Printer, Language PCL3
Date Format MDY
Display Viewing Angle 4
External Amplifier Gain 0 dBm
GPIB Address 18
Input Impedance
50 (Ohm)
1
Instrument Color Palette Default palette
Manual Tracking Adjustment 2048
Page Size Letter
Power On Last/Preset Preset
Preset Factory/User Mode
Print Format Print screen mode
Print Orientation Portrait
Printer Selection Auto
Prints/ Page 1
RS-232 Baud Rate 9600
RS-232 DTR Control On (hardware pacing)
RS-232 Receive Pacing None (data pacing)
RS-232 RTS Control IBFULL (hardware pacing)
RS-232 Transmit Pacing None (data pacing)
Speaker Volume 0
Time Date Display On
Verbose SCPI Off
1. This value may be 75 ohms, depending upon the analyzer hardware config-
uration.

Table of Contents

Other manuals for Agilent Technologies E4402B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4402B and is the answer not in the manual?

Agilent Technologies E4402B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4402B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals