Chapter 4 237
Instrument Functions: P - Z
SWEEP
Instrument Functions: P - Z
Remote Command:
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive?
Example: SWE:EGAT:POL NEG
SWE:EGAT:POL?
4.8.5.3 Delay
Controls the length of time from the time the gate condition is satisfied until the gate is
enabled.
Key Path:
Sweep, Gate Setup
State Saved: Saved in instrument state
Factory Preset: 57.7 microsecond
Resolution: 100 ns
Range: 0.0
µs to 100 seconds
Remote Command:
[:SENSe]:SWEep:EGATe:DELay <time>
[:SENSe]:SWEep:EGATe:DELay?
Example: SWE:EGAT:DEL 100US
SWE:EGAT:DEL?
4.8.5.4 Length
Controls the length of time that the gate is enabled.
Key Path:
Sweep, Gate Setup
State Saved: Saved in instrument state
Factory Preset: 461.6
µs
Range: 10
µs to 0.5 s
Remote Command:
[:SENSe]:SWEep:EGATe:LENGth <time>
[:SENSe]:SWEep:EGATe:LENGth?
Example: SWE:EGAT:LENG 1ms
SWE:EGAT:LENG?

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals