254 Chapter 4
Instrument Functions: P - Z
System
Instrument Functions: P - Z
Remote Command:
:CALibration:RF
:CALibration:RF?
Remote Command Notes: The query performs the alignment and returns a zero if the
alignment is successful.
Example: CAL:RF?
4.9.4.4.2 Align IF
Initiates an alignment on the IF assembly.
Key Path: System, Alignments, Align Subsys
Annunciation/
Annotation: Will display a series of pop-up message boxes indicating alignment
progress.
SCPI Status Bits/
OPC Dependencies: Status questionable calibration Bit 4 will be set if
Align RF fails.
Remote Command:
:CALibration:IF
:CALibration:IF?
Remote Command Notes: The query performs the alignment and returns a zero if the
alignment is successful.
Example: CAL:IF?
4.9.4.4.3 Align ADC
Initiates an alignment on the ADC circuitry.
Key Path: System, Alignments, Align Subsys
Annunciation/
Annotation: Message boxes indicating alignment of ADC progress.
SCPI Status Bits/
OPC Dependencies: Status questionable calibration Bit 6 will be set if
Align ADC fails.

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals