Chapter 4 255
Instrument Functions: P - Z
System
Instrument Functions: P - Z
Remote Command:
:CALibration:ADC
:CALibration:ADC?
Remote Command Notes: The query performs the alignment and returns a zero if the
alignment is successful.
Example: CAL:ADC?
4.9.4.4.4 Align Current IF Flatness
Initiates an alignment of the current IF flatness, for the purpose of improving absolute
amplitude within FFT Sweeps and improving group delay in some digital demodulation
measurements.
Key Path: System, Alignments, Align Subsys
Annunciation/
Annotation:
Will display a series of pop-up message boxes explaining what is being aligned (a
subset of the Align All Now message boxes).
Remote Command:
:CALibration:FLATness:IF
:CALibration:FLATness:IF?
Remote Command Notes: The query performs the alignment and returns a zero if the
alignment is successful.
Example: CAL:FLAT:IF?
4.9.4.4.5 Align Current SysGain
Initiates a fine-tuning adjustment of the system gain, primarily to correct for small
amplitude variations that occur as resolution BW is switched.
Key Path: System, Alignments, Align Subsys
Annunciation/
Annotation: Will display a series of pop-up message boxes explaining what is being
aligned (a subset of the Align All Now message boxes).

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals