5.4.2.2. TRL Calibration
1. Press Channel Next key to select channel 2.
2. Press Cal key.
3. Click Cal kit and select Cal kit which you previously defined.
4. Click Calibrate > 4-Port TRL Cal.
5. Click Thru/Line.
a) Connect “Thru” standard between Port 1 and Port 2.
b) Click 1-2 Thru/Line.
c) Connect “Thru” standard between Port 1 and Port 3.
d) Click 1-3 Thru/Line.
e) Connect “Thru” standard between Port 3 and Port 4.
f) Click 3-4 Thru/Line.
g) Click Return.
6. Click Reflect.
a) Connect “Short” or “Open” standard defined at subclass setting to Port 1.
b) Click Port1 Reflect.
c) Connect “Short” or “Open” standard defined at subclass setting to Port 2.
d) Click Port2 Reflect.
e) Connect “Short” or “Open” standard defined at subclass setting to Port 3.
f) Click Port3 Reflect.
g) Connect “Short” or “Open” standard defined at subclass setting to Port 4.
h) Click Port4 Reflect.
i) Click Return.
7. Click Line/Match
a) Click 1-2 Line/Match.
b) Connect “Load” standard between Port 1 and Port 2.
c) Click Line/Match 1[Load].
d) Connect “Line1” standard between Port 1 and Port 2.