Agilent MOI for DisplayPort Cable & Connector Compliance Tests
15
5.4.2.2. TRL Calibration
1. Press Channel Next key to select channel 2.
2. Press Cal key.
3. Click Cal kit and select Cal kit which you previously defined.
4. Click Calibrate > 4-Port TRL Cal.
5. Click Thru/Line.
a) Connect Thru standard between Port 1 and Port 2.
b) Click 1-2 Thru/Line.
c) Connect Thru standard between Port 1 and Port 3.
d) Click 1-3 Thru/Line.
e) Connect Thru standard between Port 3 and Port 4.
f) Click 3-4 Thru/Line.
g) Click Return.
6. Click Reflect.
a) Connect Short or Open standard defined at subclass setting to Port 1.
b) Click Port1 Reflect.
c) Connect Short or Open standard defined at subclass setting to Port 2.
d) Click Port2 Reflect.
e) Connect Short or Open standard defined at subclass setting to Port 3.
f) Click Port3 Reflect.
g) Connect Short or Open standard defined at subclass setting to Port 4.
h) Click Port4 Reflect.
i) Click Return.
7. Click Line/Match
a) Click 1-2 Line/Match.
b) Connect Load standard between Port 1 and Port 2.
c) Click Line/Match 1[Load].
d) Connect “Line1 standard between Port 1 and Port 2.

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals