Agilent MOI for DisplayPort Cable & Connector Compliance Tests
37
6.3.4. Return Loss Measurement (Normative)
1. Select Trace 2.
2. Press Analysis.
3. Click Fixture Simulator > BalUn and turn it ON.
4. Click Measurement > Sdd11.
5. Press Scale > Set Divisions to 10.
6. Set Scale/Div to 5 dB/div.
7. Set Reference position to 10 Div.
8. Set Reference Value to 0 dB.
9. Select Trace 6.
10. Press Analysis.
11. Click Fixture Simulator > BalUn and turn it ON.
12. Click Measurement > Sdd22.
13. Repeat from step 5 to step 8.
6.3.5. Insertion Loss Measurement (Normative)
1. Select Trace 3.
2. Press Analysis.
3. Click Fixture Simulator > BalUn and turn it ON.
4. Click Measurement > Sdd21.
5. Press Scale > Set Divisions to 10.
6. Set Scale/Div to 5 dB/div.
7. Set Reference position to 10 Div.
8. Set Reference Value to 0 dB.
6.3.6. Near End Noise (NEN) Measurement
1. Select Trace 4.