Agilent MOI for DisplayPort Cable & Connector Compliance Tests
28
6. Appendix
6.1. Defining a calibration Kit
To change the definition of a calibration kit, follow the procedure below.
1. Press Cal key.
2. Click Cal Kit > User
3. Click Modify Kit > Label Kit [User], then type in a name you want.
4. Click Define STDs >
a) 1.No Name >
Label : ”Thru”
1. STD Type : Delay/Thru
2. Offset Delay : Value defined by the fixture
3. Offset Z0 : Value defined by the fixture
4. Offset Loss : Value defined by the fixture
5. Min. Frequency : Value defined by the fixture
6. Max. Frequency : Value defined by the fixture
7. Return
b) 2. No Name >
1. Label : ”Short”
2. STD Type : Short
3. Offset Delay : Value defined by the fixture
4. Offset Z0 : Value defined by the fixture
5. Offset Loss : Value defined by the fixture
6. Min. Frequency : Value defined by the fixture
7. Max. Frequency : Value defined by the fixture
8. Return
c) 3. No Name >
1. Label : ”Open”
2. STD Type : Open