EasyManua.ls Logo

Agilent Technologies E5071C

Agilent Technologies E5071C
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent MOI for DisplayPort Cable & Connector Compliance Tests
28
6. Appendix
6.1. Defining a calibration Kit
To change the definition of a calibration kit, follow the procedure below.
1. Press Cal key.
2. Click Cal Kit > User
3. Click Modify Kit > Label Kit [User], then type in a name you want.
4. Click Define STDs >
a) 1.No Name >
Label : ”Thru”
1. STD Type : Delay/Thru
2. Offset Delay : Value defined by the fixture
3. Offset Z0 : Value defined by the fixture
4. Offset Loss : Value defined by the fixture
5. Min. Frequency : Value defined by the fixture
6. Max. Frequency : Value defined by the fixture
7. Return
b) 2. No Name >
1. Label : ”Short
2. STD Type : Short
3. Offset Delay : Value defined by the fixture
4. Offset Z0 : Value defined by the fixture
5. Offset Loss : Value defined by the fixture
6. Min. Frequency : Value defined by the fixture
7. Max. Frequency : Value defined by the fixture
8. Return
c) 3. No Name >
1. Label : ”Open
2. STD Type : Open

Other manuals for Agilent Technologies E5071C

Related product manuals