2. Purpose
This test procedure was written to explain how to use the Agilent E5071C ENA Option
TDR to make the measurements required per VESA DisplayPort Standard Version 1,
Revision 2a.
3. References
VESA DisplayPort Standard Version 1, Revision 2a
VESA DisplayPort PHY Compliance Test Specification Version 1.2b
4. Resource Requirements
1. E5071C Network Analyzer with option TDR and option 4D5 or 4K5
2. Display Port test fixtures. BitifEye BIT-1050-0001-0, or an equivalent set of fixtures
and standards.
3. Four 3.5 mm cables 20 GHz bandwidth or equivalent
(Cables of equal length and characteristics must be used for all test ports)
4. 50 Ohm terminators to terminate unused channels (ex. Agilent 909D-301)