Agilent MOI for DisplayPort Cable & Connector Compliance Tests
16
e) Click Line/Match 2[Line1].
f) Connect “Line2” standard between Port 1 and Port 2.
g) Click Line/Match 3[Line2].
h) Connect “Line3” standard between Port 1 and Port 2.
i) Click Line/Match 4[Line3].
j) Connect “Line4” standard between Port 1 and Port 2.
k) Click Line/Match 4[Line4].
l) Click Return.
m) Click 1-3 Line/Match and repeat step b) to l).
n) Click 3-4 Line/Match and repeat step b) to l).
o) Click Return.
8. Click Done to finish TRL 4-port calibration. At this point, the calibration coefficient is
calculated and saved. The error correction function is automatically turned on.
Note: Refer to “4-port TRL Calibration” in ENA online help for the detail.
5.4.3. Set System Impedance (Optional)
In case that Load standard is not 50 ohm, system impedance must be set to the actual Load
impedance.
1. Press Cal > Set Set Z0 to actual impedance of the Load standard.
2. Click Return.