Agilent MOI for DisplayPort Cable & Connector Compliance Tests
38
2. Press Analysis.
3. Click Fixture Simulator > BalUn and turn it ON.
4. Click Measurement > Sdd21.
5. Press Scale > Set Divisions to 10.
6. Set Scale/Div to 5 dB/div.
7. Set Reference position to 10 Div.
8. Set Reference Value to 0 dB.
6.4. Limit Test Settings
6.4.1. Displaying Judgment Result of Test
If a channel has a judgment result of fail, the fail message appears on the screen. It will be
judged as failed if one or more unsatisfactory trace exists within the channel.
Follow the procedure below.
1. Press Analysis > Limit Test > Fail Sign to switch the fail sign ON/OFF.
6.4.2. Setting the Warning Beeper
Beep sound that occurs when the judgment result is fail.
Follow the procedure below.
1. Press System > Misc Setup > Beeper > Beep Warning to switch the warning beeper
ON/OFF.
6.4.3. Defining the Limit Line
Set limit lines to perform pass/fail tests on the following measurement items.
1. Bulk Cable and Connector Impedance (Trace 1, 5 in Channel 1)
2. Insertion Loss (Trace 3 in Channel 2)
3. Return Loss (Trace 2, 6 in Channel 2)
4. Near End Noise (Trace 4 in Channel 2)
5. Far End Noise (Trace 1, 5 in Channel 2)