184 Appendix C
COM Object Reference
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP
Object type Property
Syntax SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP = Va lu e
Val ue = SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP
Description For the active trace of channels 1 to 4 (Ch), sets the span value used for the transformation
function of the fault location function.
Variable
For information on the variable (Ch), see Table C-2, “Variable (Ch),” on page 159.
Examples Dim Span As Double
SCPI.CALCulate(1).PARameter(1).SELect
SCPI.CALCulate(1).SELected.TRANsform.TIME.STOP = 2E-8
Cent = SCPI.CALCulate(1).SELected.TRANsform.TIME.STOP
Related objects SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt on page 180
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe on page 181
Equivalent key [Analysis] - Fault Location - Stop
Value
Description Stop value
Data type Double precision floating point type (Double)
Range Varies depending on the frequency span and the number of points.
Preset value 1E-8
Unit s (second)
Note If the specified variable is out of the allowable setup range, the minimum value (if the lower
limit of the range is not reached) or the maximum value (if the upper limit of the range is
exceeded) is set.