Appendix B 143
SCPI Command Reference
:CALC{1-4}:SRL
SRL commands
This section describes the commands specific to the SRL function E5061A/E5062A.
:CALC{1-4}:SRL
Syntax :CALCulate{[1]|2|3|4}:SRL[:STATe] {ON|OFF|1|0}
:CALCulate{[1]|2|3|4}:SRL[:STATe]?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4), turn ON/OFF SRL
measurement function.
Parameters
Query response {1|0}<newline><^END>
Example of use 10 OUTPUT 717;":CALC1:SRL ON"
20 OUTPUT 717;":CALC1:SRL?"
30 ENTER 717;A
Related commands
Equivalent key [Analysis] - SRL - SRL
Description
ON or 1 Turns ON the srtuctural return loss function.
OFF or 0 (preset value) Turns OFF the srtuctural return loss function.