4-4
Assembly Replacement
Instrument Cover
Instrument Cover
Tools Required
T-15 TORX screwdriver
T-20 TORX screwdriver
Removal
1. Remove the two strap handles (item 1) from each side of the signal generator by loosening the two screws
(item 2) on each handle.
2. Remove the four bottom feet (item 3).
3. Remove the four rear feet (item 4) from the signal generato
r by removing the four screws (item 5) that secure them.
4. Slide the i
nstrument cover (item 6) off the back of the signal generator.
Replacement
1. Reverse the removal procedure and do the following:
Torque the rear feet screws (item 4) to 21 in-lbs.
Torque the strap handle screws (item 2) to 21 in-lbs.