4-56
Assembly Replacement
A16 Line Module
A16 Line Module
Tools Required
T-10 TORX screwdriver
5.5-mm wrench or nut driver
Removal
1. Remove the rear panel assembly. (Refer to "Rear Panel" in this chapter.)
2. Remove the nut (item 1) that attaches the line module’s ground wire (A16W2) to the inside of rear panel.
3. Remove the two screws (item 2) that attach the line module (A16) to the rear panel.
4. Remove the line module from the opening in the rear panel.
Replacement
1. Reverse the removal procedure and do the following:
Torque the 5.5-mm nut that attaches the ground wire to the rear panel to 9 in-lbs.
Torque all T-10 TORX screws to 9 in-lbs.